Name: must match part name on model Content
Sterilization Pan 1

External Inspection. Blah blah. The text info for scene 1 can be set by SME in table

Sterilization Pan 2

Open the pan and inspect the contents. The text info for scene 2 can be set by SME in table

Info prompt

Select all unsterile items, if any.

Show answers prompt

Hover over the icons to see the rationale of why the item is unsterile

Blue Filter Paper

Blue filter paper present.

Blue Filter Paper - ERROR

Blue filter paper is missing and therefore the pan is unsterile.

External Sterility Indicator - ERROR

The colour change is not evident on the external indicator showing that the item was not exposed to the sterilisation process.

Tamper-proof Lock L - ERROR

Left Tamper-proof lock is missing; therefore, the item must be considered unsterile.

Tamper-proof Lock R - ERROR

Right Tamper-proof lock is missing; therefore, the item must be considered unsterile.

Internal Sterility Indicator - ERROR

The blue bar does not enter the ‘accept’ area and therefore indicates that the item did not meet the critical parameters of sterilisation.

Curved Mayo Scissors - ERROR

Debris is evident on the scissors; therefore, the entire pan is not sterile.